Electricity basics


basic electricity anaesthesia


Menu


[ Go to Home Page ]


Why should you know about electricity ?

Electricity is everywhere. There are billions of electrons flowing all around you, in operating room lights, anesthesia monitors , diathermy , ventilator controls, syringe pumps, the computer screen that you are reading now and so on and so on .....

The amazing image below shows a picture of earth taken at night from outer space. The street and house lights illuminate the countries and give an idea how widespread the use of electricity is all over the world. In the image below, how lit up is your country ?

sattelite map showing lights all over the world

Learning a bit about electricity is useful because:


[ Go to Top of Page ] [ Go to Home Page ]


Electrical current and ampere

One often hears the words "current", "voltage" and "resistance". It is very important to understand these three concepts well. Let us start by explaining what electric current is. Let us imagine that an electrical current is going through the light brown wire below.

what is currant

When we say that an electrical "current" is flowing, what is it that is actually flowing ? Current is the flow of electrons. Electrons can be thought of as negatively charged "particles". The movement of these electrons is called current.

electrons going in wire

Current is measured in units called 'amperes'. The number of amperes in a wire relates to how many electrons pass a cross section of the wire per second.

current

A current of one ampere relates to a certain number of electrons passing a cross section of the wire in one second. This number is absolutely huge, and please don't bother to remember it. The wire below is carrying one ampere of current.

one ampere of electrons

Higher the current (higher amperes) higher is the number of electrons that pass a cross section of the wire per second.

low and high current

In reality, it is really tough to count zillions of electrons.

zillions of electrons

Since it is so difficult to count so many electrons, the official definition ("International System of Units") of the ampere relies on a more "practically measurable definition". To you and me, even this definition would still appear very impractical, but then we aren't physicists (luckily). I have mentioned this 'practical' definition below. Please don't memorize it, I have put it here only for completeness sake.

SI definiton of ampere

So in summary, ampere relates to the amount of electrons flowing across a wire per second.


[ Go to Top of Page ] [ Go to Home Page ]


Potential difference and volts

Let us imagine that you want to install a beautiful waterfall in your garden (also imagine that you are rich enough to have a big garden).

waterfall

To save money, you decide to build what you think is a waterfall. It looks like this.

useless waterfall

Unfortunately the water doesn't flow at all. You cannot understand why.

why isn't the water flowing

The reason why there is no water flow is that there is no energy difference in the flat pond. For water to flow, there needs to be a difference in energy between one area and another. Then the water will flow from the area of high energy to the area with a low energy. In the case of a waterfall, this energy can be given by elevating one part of the water. The elevated water has more energy (due to gravity) than the water at the base, making water flow.

higher energy to lower energy water flow

As you admire your waterfall and show it off to your friends, you suddenly discover that it has stopped flowing! The reason is that the higher water tank is now empty !

no more water

So you decide that the solution is to reuse the water and you install a water pump. The pump takes the water from the bottom and pumps it to the upper water tank. Now you have a continuous beautiful waterfall.

water pump keeps the waterfall flowing

Let us examine this in terms of energy. In the case of the waterfall, it is gravity that determines energy. Higher the water, more is the potential energy does it have. That is why the water in the higher tank has a higher potential energy than the lower tank which has a lower potential energy. The difference in energy makes the water flow. As the water flows down the water fall, it loses potential energy. Once it reaches the bottom, it therefore has a lower potential energy than the upper tank.

higher tank has more potential energy than the lower tank

Once the top tank is empty, of course water wont flow back upwards on its own. This is because water does not flow from an area of low potential energy to an area of high potential energy.

water needs energy

For the same reason, if you simply connected a pipe from the lower tank to the upper tank, the water won't go back to the higher tank.

water potential energy

If we want our garden waterfall to keep working we need to make water return to the upper tank. For this, we need to provide energy to the system. This is possible with a water pump. Because water in the upper tank has a high energy, it flows down, giving up energy in the process to end up in the lower tank. The pump then takes the water with lower potential energy and pumps it to the upper tank, repeating the cycle over and over again. Now finally we have a beautifully working garden waterfall.

pump

Now let us move to something electrical. Instead of having a waterfall, let us imagine that we want to light up a light bulb. A light bulb is something that lights up when current passes through it.

light bulb

Let us connect up some wires to the bulb to try and make it light up. Oh no! The bulb does not light up. We can see that the wire has electrons in it, but the bulb does not light up.

dark bulb

The reason that the bulb does not light up is that there is no flow of electrons. This is similar to the water not flowing in the level pond.

Let us see the wire in close up to see why the bulb is not lighting

.

You will see that the electrons in the wire are moving randomly in all directions and are not flowing in a particular direction. You will recall that current is the flow of electrons. Since the electrons in the wire are all not flowing in a particular direction, there is no current passing through the bulb, and hence the bulb does not light up.

no net flow of electrons

This is similar to the situation where the waterfall did not work without a pump. The waterfall system has water, but the water is not moving, due to the lack of energy. Even if you connect a pipe between the tanks, water will not flow on its own to the upper tank.

waterfall needs energy

In the waterfall, you will recall that it was the pump that produced the energy.

pump provides the energy for the waterfall

Just like the water in the waterfall, electrical circuits need a potential energy difference to make the current flow.

The image below shows two wires. The upper wire has no potential difference across it. The electrons are randomly moving in all directions with no net flow. Therefore this wire is not carrying any current. The lower wire has a potential difference across A and B. It makes the electrons flow in a particular direction and therefore this wire is carrying a current.

potential difference across A and B

There are a variety of sources that can produce a potential difference that can make a current flow in an electrical circuit. You can think of a 'potential difference source' as being like the pump in the waterfall we discussed. The pump makes water flow around the system. In an electrical circuit, the potential difference source makes the electrons (i.e. current) flow around the circuit.

potential difference source

Potential difference sources have a ' negative pole' and a 'positive pole'. Current leaves the negative pole and returns to the positive pole.

positive and negative terminals

As mentioned before, there are many types of potential difference sources, and many of these will be discussed later. For now, let us discuss a common potential difference source that you may have used at home and in the hospital. This source is the common disposable battery. You may have put these in the clocks in your home. Most laryngoscopes use these batteries to power the light bulb.

battery

Let us now connect this battery to the bulb we were trying to light up before. The potential difference generated in the battery makes the electrons move. The current goes from the negative pole, through wires to the bulb, goes through the bulb and makes it light up, and returns by wire to the positive pole.

battery connected to a bulb

The unit of measurement for potential difference is "volt". More the potential difference across the poles, more is the volts.

A typical house hold battery has 1.5 volts, or abbreviated as “1.5 V”

battery with potential difference of 1.5 volts

On the other hand, potential difference coming out of wall sockets can be quite high (e.g. 120 volts or 230 volts).

United Kingdom power outlet

Now I must confess that drawing all this stuff can be quite tiring. Fortunately, when discussing electricity, one doesn't have to draw the "real" thing all the time. Instead, one can use symbols. For an example, the symbol for a battery is:

battery symbol

The shorter vertical line represents the negative pole.

battery symbol

If you really want to remember which is negative and which is positive, here is a simple memory aide. Just remember ' n ' is shorter than ' p '

battery  symbol memory aid

Similarly, a light bulb can be represented by a symbol.

bulb symbol

Electrical circuits can be drawn much more easily than using realistic diagrams. For an example, this circuit ....

realisitic circuit diagram

... can be easily drawn as ....

battery lamp circuit

In practice, real medical equipment circuits can be very complex and the use of symbols becomes very convenient. Below is a very simple (for engineers) electrical circuit drawn using symbols.

complex circuit

Relationship between potential difference and current.

Now let us return to potential difference and current. Recall that current is the flow of electrons. Potential difference is what makes the current flow. In our waterfall example, we said that potential difference was like the difference in height of the upper and lower tank, and the difference made the water flow.

water height detrmines the flow

Take the two waterfalls below. Waterfall A has a large difference in height whereas waterfall B has small difference in height. Fall A has a higher energy difference than fall B. Consequently the flow rate of water in fall A will be much more than the flow rate of water in fall B. In other words, water flow rate is proportional to the energy difference.

different heights of waterfall

This is the same in electrical circuits. If you increase the potential difference, more electrons will flow. Take the example below. In the top circuit, the potential difference is 1.5 volts and this results in a certain current flow. In the bottom circuit, the potential difference is doubled to 3 volts. This doubles the current flow, making the bulb light up brighter.

higher the potential difference, more is the flow

There is a famous electrical law called “Ohms Law”. While Ohms Law will be explained in a bit more detail later, what we have already discussed forms “PART” of the law.

I.e. Part of Ohms Law states that “for a given wire (conductor), current flow is directly proportional to the potential difference across it.”

part of ohms law

Sources of potential difference

There are many sources of potential difference.

Single use batteries :

So far in our discussions we have used batteries as our source of potential difference. We will discuss batteries first and then other sources of potential difference.

battery

Batteries are used frequency at home to power small items such as clocks. In hospital, laryngoscope bulbs are typically powered by batteries. Batteries come in all sorts of shapes and sizes.

batteries

A chemical reaction in the battery generates the potential difference. Once the chemical reaction has "finished", the battery no longer generates a useful potential difference and the battery is thrown away.

primary cell

Rechargeable batteries :

These batteries have chemical reactions that are "reversible". If you apply a potential difference and current to a rechargeable battery, chemical reactions in the battery "store" this energy and release it when required. Rechargeable batteries are very frequently used in household products such as mobile phones and laptops. They are also very frequently used in many hospital devices such as syringe pumps, portable monitors, motorised operating tables etc. It is extremely important to keep devices using rechargeable batteries "charged". If you don't keep things correctly charged (e.g. forgetting to plug the device to a wall power outlet) the batteries may run out early and unexpectedly shut down a piece of critical equipment, such as a syringe pump during a patient transfer.

battery dead

Electrical Generator:

Power to our homes and hospital comes from huge electrical generators. These are devices that convert rotating motion to electricity.

generator

Water in a dam can be used to turn the generator (hydroelectric power).

hydroelectric power generation

Many power stations use steam to turn the generator. The steam is made by heating water, which can be done by energy sources such as oil, gas, coal, or nuclear power.

electricity generation

Many power stations are connected to each other by a network of wires called the “electrical grid”. The grid allows the system to share resources and provide backup power if one power station fails. The grid distributes power to hospitals, homes and other consumers.

grid

The grid is carefully controlled to meet the electricity demands of the consumers. For an example, grid engineers study television programs closely because they know that after popular television programs end (e.g. after an important sporting event) , millions of people get up and heat water in electric kettles for coffee or tea. For this, grid engineers rapidly activate special power stations that can boost electricity supply to the grid quickly to meet this demand. For example, during the Royal Wedding in the United Kingdom, the electricity demand increased by 2,400,000 watts which is the equivalent of nearly a million kettles being turned on at the same time.

Royal Wedding

Emergency Generator:

Like everyone else, hospitals also get their electrical supply from the electrical grid. If the grid fails to deliver power, critical patient support equipment such as monitors, ventilators, anesthetic machines, etc. may stop functioning. For this reason, hospitals have their own emergency power supplies. For brief periods of power failure, huge rechargeable batteries are able to supply the necessary power. For longer periods, the hospital will have a generator which is usually turned by a diesel oil engine. This generator comes on automatically when a power failure is detected.

diesel backup generator

Concept of Resistance

In the world of electricity, there are three friends that are always present together and affect each other.

three friends

We have already discussed two of them, i.e. current (amperes) and potential difference (volts). Now let us discuss resistance, the third friend.

In a circuit, electrical resistance is something that “resists” the flow of electrons (i.e. resists the flow of current). Let us go back to our waterfall diagram. However, we now replace the waterfall with a pipe that connects the two tanks. The water flows in pipe from the upper tank to the lower tank due the difference in potential energy.

pipe waterflow

Let us now take two such arrangements. In both , the water tank height is the same. I. e, both systems have the same energy. However, arrangement A has a narrow pipe and arrangement B has a wide pipe.

resistance

The flow rate of water in pipe A is slower than the flow rate of water in pipe B. This is because pipe A, being narrower has a higher resistance to water flow than the wider pipe B.

pipes

The flow rate and resistance have an "opposite" relationship. When one increases, the other decreases. In more technical language, we can say that the flow rate is inversely proportional to resistance.

water flow equation

Electricity is similar. Let us take a wire and light bulb connected to a potential difference of 1.5 volts. This potential difference makes current flow in the wires and light bulb.

high resistance

Suppose, keeping the same potential difference, we use a wire with a low resistance (thicker wire below), the current flow will now be easier and more.

high an dlow resistance

In other words, high resistance results in a low current flow. A low resistance results in a high current flow. We can put this into an equation where current flow is inversely proportional to the resistance of the wire it is traveling in.

current inversely proportional to resistance

Ohm's Law

A while ago we saw part of Ohms Law which stated that the current flow across a wire is directly proportional to the potential difference across that wire.

current flow is proportional to potential difference

And more recently we leant that current flow is inversely proportional to the resistance.

current is inversely proportional to resistance

Now you are ready for the “full” Ohms Law ! Let us combine the two relationships you just learnt.

Ohms Law states that: Current flow through a conductor (wire) between two points is directly proportional to the potential difference across the two points, and inversely proportional to the resistance between them. The Ohms Law can be shown by the equation in the yellow box.

Ohms Law.

So you can see that the three friends, volt (potential difference), resistance and current are best of friends. And just like friends, if you get two together, they will talk about the third friend.

friends

Similarly, by rearranging the Ohms Law equation, if you know two of the values, you can find out what the third one is.

Ohms Law

It is time to use the international abbreviations in the equations. Note that the international abbreviation of Current is not " C ". Instead , it is " I " (capitol i ). This bizarre abbreviation has to do with history, where current used to represented in the French language.

abbreviations

Now here are the equations we saw before, using the abbreviations.

equations

You don't need to memorize all the above three equations. Just remember one, and you can work out the rest when needed.

V equals IR

Ohms Law is a very important law in the world of electricity. It was discovered by Mr. Ohm.

Mr. Ohm

The unit of resistance is named after him. The unit of resistance is called "Ohm" and its symbol is shown below.

Ohm

Let us take a example of Ohms Law in clinical practice.

Let us imagine that you have attached ECG leads to the patient. You then plug the ECG lead plug to the monitor.

Unfortunately, as you are plugging the two plugs together, a piece of dirt gets into the plug. This makes the yellow wire have a poor contact with its counter part, increasing its resistance.

dirty plug

According to Ohms Law, the high resistance affects the voltage and current in the wire, disturbing the ECG tracing.

ECG poor connection

The plug is unplugged and the dirt removed. Now the contact is good and the resistance is now low.

good connection

Low resistance makes the current flow easily and you get a much better ECG tracing.

good tracing

DC and AC current

You will recall that current is the flow of electrons.

electron flow is current

The electrons (i.e. current) flows from the "negative" pole to "positive" pole. Current therefore has a "direction" of flow.

direction of current

The direction of the current will change if the negative and positive poles of the source exchange.

polarity change

There are two main types of current flow : "Direct Current (DC) and Alternating Current (AC).

Basically, in DC, the "direction" of current flow remains constant over time, whereas, in AC, the direction of flow of current keeps alternating from one direction to the other. Let me try and make things clearer for you.

To make the diagrams a little less crowded, I have simplifying them a bit. The potential difference source is represented with a minus and plus sign. The green arrow will show the direction of current. Remember, current always flows from negative to positive.

potential difference

I will now describe to you the difference between DC and AC. Let us start with DC. Below is a series of images showing a circuit with a DC power source. The repeated images show you what happens over a time period. You will notice that over time, that the current has NOT changed direction (green arrow remains in the same direction). This is the fundamental property of DC: the current does not change direction.

DC

AC is quite weird when compared to DC current. However, there is a good reason why engineers use AC in some situations and this will be explained later. In AC, the negative pole and positive pole exchange themselves repeatedly.

AC

Since current travels from negative to positive, this means that the current also changes direction repeatedly.

AC changes direction

The series of images below show AC over time. The positive pole and negative pole keep alternating. In response to this, the direction of current also keeps changing repeatedly.:

AC

AC

This repeated alternating of current direction happens quite fast. In many countries , it happens 50 or 60 times a second (i.e. frequency of 50 Hz or 60 Hz). In the USA, it happens 60 times a second (60 Hz). In the United Kingdom , India and Sri Lanka, it happens at 50 Hz. The animation below may help you to appreciate how rapid this is. The negative and positive poles alternate 50 times second. If you suffer from photosensitive epilepsy, please skip this animation.

AC 50 Hz

So far I have shown AC current where the polarity changes suddenly from to the other. In reality, the change happens is a 'smoother way' as will be discussed next.

It can be quite difficult to visualise the changes that happen with AC. Fortunately there is an instrument called an "oscilloscope" that can help visualize what is happening with AC. An oscilloscope shows voltage changes over a time period. The vertical axis represents voltage and the horizontal axis represents time.

oscilloscope

If an "oscilloscope" sounds complicated, don't worry because I am sure you have used one everyday in your professional life. An ECG (EKG) monitor is a type of oscilloscope. It measures the potential difference of the heart over time.

ECG monitor

Below is our simplified oscilloscope. RIght now, nothing is connected to the oscilloscope probes (black and red probes) and therefore the the tracing (blue line) remains on the baseline.

oscilloscope

Let us first examine DC with the oscilloscope. We connect a battery ( a good source of DC) across the measuring probes. The oscilloscope shows a straight line reading (blue line) above the baseline (broken line).

oscilloscope

Oscilloscopes show change in direction of current flow (i.e. changing the plus and minus) by the trace crossing the baseline. In this example, the trace crosses the baseline from the top to the bottom.

polarity change

Changing the polarity again makes the tracing cross the baseline again. This time it crosses the baseline from the bottom to the top.

The above was shown to you just to make it clear how change in polarity makes the trace cross the baseline. When we connect a DC source such as a battery, there is no change in polarity. Therefore the tracing does not cross the baseline.

DC

Now let us examine AC with the oscilloscope. You know that, unlike DC, the polarity of AC keeps changing. So you would expect to see something like this.

square wave

However, the above square waveform is NOT what you see. Instead you will see something much more graceful.

ballet dancers

The actual AC oscilloscope tracing you see is a graceful curved waveform that crosses the baseline repeatedly. The repeated crossing of the baseline indicates that in AC the polarity keeps changing.

AC

The reason why you see the graceful curves instead of the sudden square changes is because the polarity in AC does NOT change suddenly as shown below.

AC polarity does not change suddenly

Instead, the change in AC happens "gently". The potential first starts to decrease and eventually becomes zero. Then the potential difference starts rise again, but with the polarity reversed. The potential difference keeps rising till it reaches the maximum. The changes in polarity continue in this fashion.

gentle polarity change

The following series of oscilloscope images graphically show the generation of the AC waveform.

AC waveform

AC waveform

Note that AC has periods where there is no current flow.

AC waveform

The change of polarity causes the tracing to cross the baseline.

AC waveform

AC waveform

AC waveform

AC waveform

And the cycle repeats itself.

You would have noticed that AC has periods where there is zero potential difference. During these times the current flow is zero.

zero points

Normally one does not notice these zero periods because they happen so fast. However, sometimes you can see these off periods in fluorescent lights (tube lights). The off periods may make the light flicker (faster than what is shown below, the computer animation is not fast enough).

flickering tube light

In physics, the curved waveform of AC can be described as being a "sine wave". The mathematics of such a wave are described by a "sin" function in the complicated equation shown below, which I hope you will not memorize.

sinewave

Below are commonly used symbols to represent AC and DC power supplies.

AC DC symbols

If you look around you will discover that your world is full of symbols. When you are bored have a look around your operating room and you will find plenty of hidden symbols. Typically they will be found on electrical equipment. Look where monitoring leads and power leads connect.

symbols

Just to recap, the image below shows a DC waveform and an AC waveform together for comparison.

DC and AC

Why are our homes and hospitals supplied with AC and not DC ?

The electrical current that comes out of the wall sockets in homes and hospitals is mostly AC. AC looks more complicated than DC, so why do they use AC to power our hospitals and homes ?

To explain why AC and not DC is supplied to your hospital and home, you need to understand how a device called a "transformer" works. A transformer "transforms" voltage to a higher voltage or a lower voltage. If it transforms the input voltage to a higher output voltage, it is called a "step up" transformer.

step up transformer

If it transforms the input voltage to a lower output voltage,it is called a "step down" transformer.

step down transformer

Let us see how a transformer works. The transformer uses two very important properties in the world of electricity. One is called "electromagnetic induction" and this is described first. Let us take a coil of wire.

secondary coil

Electromagnetic induction refers to a phenomenon where if a wire (or coil of wire as in our example) is exposed to a CHANGING magnetic field, a current will be induced in the wire. In the example below, the changing (i.e. repeatedly increasing and decreasing) magnetic field is represented with the blue arc with arrows.

It is important that the magnetic field is changing. A non changing magnetic field will NOT induce a current in the coil.

secondary

The second phenomenon used in transformers is that when a wire carries an electric current, it generates a magnetic field.

In the example below, the wire (coil) is carrying alternating current (AC). Since the current is changing, it produces a changing magnetic field (shown by blue arc with arrows).

If we use DC instead of AC in the wire, the magnetic field produced will be non changing since DC is non changing.

Now we can explain how the transformer works. The input AC goes into the primary coil (pink). This produces a changing magnetic field (blue arc with arrows). The changing magnetic field induces a current in the secondary coil (green). In this way, electric energy is transferred from the primary coil to the secondary coil.

transformer

At this point you can see that a transformer works only with AC, because it needs a changing magnetic field to transfer energy across it. If you used DC, the transformer would not work. The magnetic field would be non changing and thus would not transfer energy across to the secondary coil.

DC transformer would not work

Whether the transformer behaves as a step up or a step down transformer depends on the ratio of the loops in the primary coil and secondary coil. In a step up transformer, the secondary coil has more turns than the primary coil. Similarly, in a step down transformer, the secondary coil has less turns than the primary coil.

transformer

Now we understand that transformers can transform the voltage up or down. We also understand that we need AC for transformers to work. Now the question is, why are transformers so important ?

It has got to do with the transmission of electricity. Earlier on,we discussed how electricity is generated at the power station.

power station

This power station may be many hundred miles / kilometers from your home or hospital and this means that the electricity has to travel very far before reaching you. One big headache for the power company is that when electricity travels in wires, it loses energy. If this happens over huge distances, there will be nothing left when the wire reaches you.

There is a physics principle that wires carrying a low voltage have higher losses than wires carrying a high voltage. (The explanation for this beyond the scope of this website.)

Therefore, to minimise losses, the power company transmits electricity at high voltages.

These high voltage power lines can often be seen crisscrossing the landscape. Next time you are outside, do have a look for them.

pylon

This is where transformers come in. Relatively low voltage from the generator is raised by a step up transformer to a high voltage. As the wires reach you, the high voltage is reduced using a series of step down transformers. The only practical way to generate the high voltages and subsequent reduction needed for economical power transmission is to use transformers. And that is why AC is used for power transmission, all way to the wall socket in your hospital and home.

high voltage transmission


[ Go to Top of Page ] [ Go to Home Page ]


Electrical Safety

I hope that you have now have a basic understanding of electricity. The next section I suggest that you read is "Electrical Safety". You can find it on the links on the home page or you can click HERE to go Electrical Safety.